Chevrolet Cruze manuals

Chevrolet Cruze Repair Manual: Installation Procedure

  1. Clean and prepare the attaching surfaces for welding.
  2. Apply structural adhesive to the attaching flanges.

  1. Position the body side outer lower panel on the vehicle
  2. Verify the fit of the body side outer lower panel.
  3. Clamp the body side outer lower panel into position.

  1. Spot weld accordingly.
  2. Apply the sealers and anti-corrosion materials to the repair area, as necessary. Refer to Anti-Corrosion Treatment and Repair.
  3. Paint the repaired area. Refer to Basecoat/Clearcoat Paint Systems.
  4. Install all related panels and components.
  5. Connect the negative battery cable. Refer to Battery Negative Cable Disconnection and Connection.
  6. Enable the SIR system. Refer to SIR Disabling and Enabling.
Removal Procedure
Warning: Refer to Approved Equipment for Collision Repair Warning in the Preface section. Warning: Refer to Glass and Sheet Metal Handling Warning in the Preface section. Disable the SIR Syst ...

Rear Compartment Floor Panel Sectioning (MAG-Welding)
Note: According to different corrosion warranties, only the regional mandatory joining methods are allowed. ...

Other materials:

Installation Procedure
Cut the front hinge pillar body in corresponding locations to fit the remaining original panel. The sectioning joint should be trimmed to allow a gap of one-and-one-half-times the metal thickness at the sectioning joint. Create a 50 mm (2 in) backing plate from the unused portion of ...

SIR Service Precautions
SIR Service Precautions Warning: When performing service on or near the SIR components or the SIR wiring, the SIR system must be disabled. Failure to observe the correct procedure could cause deployment of the SIR components. Serious injury can occur. Failure to observe the correct procedure ...

Brake Pedal Position Sensor Calibration
Calibration Criteria Note: Do not apply the brake pedal during the brake pedal position (BPP) sensor calibration procedure. Any movement of the brake pedal during this procedure will cause the calibration procedure to fail. If this occurs, the sensor calibration must be repeated. Brake pedal ...

© 2011-2026 Copyright www.ccruze.com